Atrium's Ocean Chest Drain Setup and Operation Part 2

  • 0:00 - 0:07
    Let's review these instructions and actually see how
    fast and efficient, Atrium's Ocean chest drains are to
  • 0:07 - 0:13
    set up and use. First, remove the CSR wrapped unit
    from the outer bag. Carefully open the CSR wrap
  • 0:13 - 0:19
    using sterile technique and enter the unit into the
    sterile field. Each unit comes equipped with a large,
  • 0:19 - 0:26
    easy to grip handle that makes hand-off and patient
    transport easier and safer than ever before.
  • 0:26 - 0:31
    The sturdy, swing-out floor stand should be opened for
    set-up convenience and for placement onto the floor.
  • 0:31 - 0:38
    To begin, fill the water seal with sterile water or sterile
    saline by the pre-attached, pre-measured funnel,
  • 0:38 - 0:45
    by holding the funnel downward and filling it to the
    top, as shown. By raising the funnel, water will quickly
  • 0:45 - 0:53
    empty into the drain and fill the water seal to the
    required 2cm level, indicated by the dotted, "fill to" line.
  • 0:53 - 0:59
    Remove the funnel and discard after use. Next,
    connect the chest drain to the patient by inserting the
  • 0:59 - 1:06
    patient tube connector directly into the thoracic
    catheter. The patient is now on "gravity drainage or
  • 1:06 - 1:13
    "to water seal." To apply suction when prescribed by a
    physician, the suction control chamber can be filled to
  • 1:13 - 1:19
    the desired pressure level simply by removing the
    tethered vent plug, pouring water to the desired
  • 1:19 - 1:24
    suction control pressure level,
    and replacing the vent plug.
  • 1:27 - 1:33
    As with any chest drainage system, the drain must be
    placed below the level of the patient's chest and kept
  • 1:33 - 1:39
    in an upright position to maximize drainage
    efficiency. We recommend hanging the unit bedside,
  • 1:39 - 1:45
    using the multi-position hangers provided. The
    integrated flexi-hangers accommodate today's newer
  • 1:45 - 1:50
    bed designs and allow single-hand placement and
    transport. When you're ready to initiate suction,
  • 1:50 - 1:57
    connect the suction source to the chest drain suction
    connector, or suction control stopcock, and slowly
  • 1:57 - 2:04
    increase suction at the suction source until constant,
    gentle bubbling occurs in the suction control chamber.
  • 2:04 - 2:11
    Available on most Atrium Ocean models is a patented
    suction control stopcock which provides simple and
  • 2:11 - 2:17
    effective control of vacuum to the drain for more
    efficient system operation, with less bubbling noise
  • 2:17 - 2:24
    and less evaporation. Simple to use and operate,
    Atrium's Suction Control Stopcock can be adjusted as
  • 2:24 - 2:30
    necessary to increase or decrease suction control
    bubbling. When applying wall suction without a wall
  • 2:30 - 2:37
    regulator, simply turn the stopcock to the off
    position prior to connecting to the suction source.
  • 2:37 - 2:43
    Once connected, use the suction control stopcock to
    slowly increase suction until you obtain a constant
  • 2:43 - 2:46
    level of gentle bubbling in
    the suction control chamber.